Cause and Treatment of Inappropriate Therapy of Implantable Cardioverter Defibrillators

Dong-mei WANG,Ya-ling HAN,Hong-yun ZANG,Hai-bo YU,Xiao-guang WU,Shou-li WANG,Zu-lu WANG,Quan-min JING,Hong REN
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2008.02.002
2008-01-01
Abstract:Objective The purpose of this study was to analyze the causes of inappropriate therapy and explore the treatment measure in patients with implantable cardioverter defibrillators (ICD). Methods Thirty four patients 24 men and 10 women, mean age (62. 54 ± 11.17) years received ICDs. Among these patients,21 received single-chamber ICD, 3 received double-chamber ICD and 10 received triple-chamber ICD (CRT-D).The mean follow-up duration were(20.72 ±16.89) months. Results Six(17.6%) patients (3 singlechamber ICD, 3 CRT-D) experienced inappropriate detections and therapies of ICD during follow-up. One patient with single-chamber ICD received an inappropriate shock due to myopotential interference (noise). The inappropriate shocks were caused by atrial fibrillation (AF) with rapid ventricular response reaching into ventricular fibrillation (VF) detect-zone in 4 patients. Device in the sixth patient with CRT-D (V350,ST. Jude Medical) detected supraventricular tachycardia (SVT) for 7 times. SVT were terminated by shocks after failed in ATPs when the lasting time of SVT was more than 60 seconds, because the longest time of discriminating SVT in device was 60 seconds. AF for 4 times in 2 patients is successfully terminated by ATP in FVT detectzone. Forty shocks 6 patients were caused by inappropriate detections, of which 14 occurred in single-chamber ICD and 26 occurred in triple-chamber ICD. The frequently shocks in a short time occurred in 6 patients. A patients received 7 shocks within 3 hours. Conclusions The rapid ventricular rate of AF in VF detect-zone was the main cause of inappropriate detections of ICD. Therefore control the ventricular rate of AF with medicine is important was patients with ICD. If the fast ventricular rate of AF is not controlled by medicine, the radiofrequency ablation of AF is the choice.
What problem does this paper attempt to address?