Inappropriate ICD Discharge Due to T-Wave Oversensing in a Patient with Short QT Syndrome

Yaxun Sun,Ping Zhang,Xuebin Li,Jihong Guo
DOI: https://doi.org/10.1111/j.1540-8159.2009.02538.x
2010-01-01
Pacing and Clinical Electrophysiology
Abstract:A 45-year-old man with diagnosis of short QT syndrome underwent successful implantation of dual-chamber implantable cardioverter-defibrillator (ICD) to prevent sudden cardiac death. Ten days after implantation, the patient was treated by inappropriate ICD discharges for frequent T-wave oversensing. After careful reprogramming of the ICD, the T-wave oversensing was eliminated and no T-wave oversensing or inappropriate discharge was documented during 6-month follow-up period. (PACE 2010; 113-116).