Inappropriate ICD discharges due to "triple counting" during normal sinus rhythm

Ejaz Khan,Apostolos Voudouris,Stephen R Shorofsky,Robert W Peters
DOI: https://doi.org/10.1007/s10840-006-9062-0
Abstract:Objective: To describe the clinical course of a patient with multiple ICD shocks in the setting of advanced renal failure and hyperkalemia. Methods: The patient was brought to the Electrophysiology Laboratory where the ICD was interrogated. Results: The patient was found to be hyperkalemic (serum potassium 7.6 mg/dl). Analysis of stored intracardiac electrograms from the ICD revealed "triple counting" (twice during his QRS complex and once during the T wave) and multiple inappropriate shocks. Correction of his electrolyte abnormality normalized his electrogram and no further ICD activations were observed. Conclusion: Electrolyte abnormalities can distort the intracardiac electrogram in patients with ICD's and these changes can lead to multiple inappropriate shocks.