Clinical safety of cardiac resynchronization therapy and implantable cardiac defibrillator implantation compatible with magnetic resonance imaging

Qiaoyuan Li,Meixiang Xiang,Yangang Su,Wenlong Dai,Dongfang He,Dongping Fang,Chengjun Guo
DOI: https://doi.org/10.3760/j.issn.1673-4777.2020.01.008
2020-01-01
Abstract:Objective To observe the safety of magnetic resonance imaging (MRI)-conditional cardiac resynchronization therapy,implantable cardiac defibrillator(CRT-D) implantation and postoperative MRI performance.Methods From December 2017 to March 2018,12 heart failure patients with poor response to medication therapy were implanted MRI-conditional CRT-D (Inventra 7 HF-T QP).Pacemaker parameters were observed during implantation,1 month after operation,before MRI scanning and 1,3 months after MRI examination.Results There were 7 males and 5 females.MRI was performed in 10 cases(83.3%).There were no significant changes of CRT-D electrical parameters during implantation,1 month after operation,before and 1,3 months after MRI scanning(all P>0.05).The variations of electrical parameters at 1 month after MRI were (51 ± 31)Ω for right atrial impedance,(116 ± 90) Ω for left ventricular impedance,(26 ± 4) Ω for right ventricular impedance and (5.6 ± 2.1)Ω for defibrillation impedance.Three cases showed slight artifacts in MRI which did not affect diagnosis,while 1 case had severe artifact around the heart that hindered image reading.No threshold increasing,sensation abnormality and lead dislodgment,myocardium perforation were observed.Conclusion MRI-conditional CRT-D pacemaker is feasible and MRI scanning can be performed safely after implantation.