The feasibility of in-vitro testing in ICD implantation procedure

Sun Yaxun,Jiang Jiangfen,Sheng Xia,Jiang Ruhong,Chen Shiquan,He Hong,Fu Guosheng,Jiang Chenyang
DOI: https://doi.org/10.3969/j.issn.1005-0272.2013.05.003
2013-01-01
Abstract:Objective To evaluate a simple clinical method to test sensing ability and avoid VO in implanting procedure. Methods In ICD replacement procedure, we connected the in-vitro testing ICD with invivo sensing lead through active fixation lead and bridging cable. The amplitude and morphology of ventricular sensing and pacing signals of different ICDs were analyzed and compared with those of in-vivo ICDs. The invitro ICDs were programmed to the most sensitive to reveal the possible VO. Results Comparing with in-vivo ICD, in-vitro ICD testing provided similar signal pattern of ventricular sensing both in sensing and pacing mode.Threshold testing showed the same results in in-vitro and in-vivo mode. Impedance of pacing of in-vitro increased slightly(~35Ω). In 8 patients, in-vitro ICDs showed no VO under the most sensitive setting. Short term follow-up showed no VO episode or inappropriate therapy. Conclusions In-vitro ICD testing could mimic invivo ICD sensing by similar amplitude and morphology of ventricular sensing and provided a simple method to reveal possible ventricular oversensing before ICD implantation.