Clinical Study of Tri-chamber Cardiac Pacemaker

郭继鸿,李学斌,许原,张海澄,金华,陈冰
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.1998.03.112
1998-01-01
Abstract:Objective:To evaluate the efficacy and safety of tri-chamber cardiac pacemaker on treating paroxysmal atrial tachyarrhythmias such as atrial tachycardia, atrial flutter and atrial fibrillation.Methods:Five patients were implanted tri-chamber pacemaker with bi-atrial synchrony. Three leads were placed into coronary sinus (CS), right ventricle (RV) and right auricle (RA) respectively via left subclavian vein. CS lead and RA lead consisted of new bi-atrial lead through a "y" shape adapter after threshold test. DDD pacing and bi-atrial AAT pacing mode were used after implantation.Results:DDD pacing shown obvious efficacy on 2 patients with sinus bradycardia and less atrial premature beats. The other 3 cases shown efficacy when changed to bi-atrial AAT mode, 2 of them improved obviously. There were no complications after 4-6 months followup in all five patients.Conclusion:Tri-chamber bi-atrial synchronized pacemaker can successfully prevent atrial tachyarrhythmias caused by intra-atrial conduction block. It is safe and has less complications in spite of some difficult implanting techniques.