The Effect of CARE- Kv on Radiation Dose and Image Quality of Dual Source Coronary CT Angiography

张琰琰,杨文洁,张欢,曹立秀,潘自来,陈克敏,严福华
DOI: https://doi.org/10.13437/j.cnki.jcr.2013.11.035
2013-01-01
Abstract:Objective To discuss effect of CARE- kv on radiation dose and image quality of dual source coronary CT angiography( CTA). Methods Consecutive patients( 393 patients) were respectively enrolled: Group A( 189 patients) underwent dual- source coronary CTA with CARE- kv and CARE Dose 4D,in whom tube potential and tube current were selected automatically; group B( 204 patients) underwent CTA with a BMI- based tube potential and CARE Dose 4D. Quantitative image quality parameters,including image noise,attenuation,CNR and SNR,the CT density in a region of interest placed in the aortic root,LM and RCA were measured. The scores of imaging quality were evaluated by two experienced reviewers and CNR,SNR,CTDIvol,DLP were compared between the two group. Results The value of CNR and SNR in group A were significantly higher than that in group B( P 0. 05). The value of CTDIvol in group A were 32. 4% significantly lower than that in group B( t = 0. 098,P = 0. 003),DLP were 22. 93% significantly lower( t = 4. 838,P = 0. 043) with a significantly more frequent use of 80 kV( χ2= 73. 289,P 0. 05). Conclusion The application of CARE - kv in dual source coronary CT angiography can significantly reduce radiation dose and improve image quality.