Study of dual-source CT chest examination using 70kv tube voltage in non-overweight patients

姜倩男,张铎,孟姮,刘怡
DOI: https://doi.org/10.3969/j.issn.1007-4287.2017.05.017
2017-01-01
Abstract:Objective To explore the value of 70kv tube voltage scan in chest CT examination of non-overweight patients.Methods Sixty patients underwent dual-source computed tomography (CT) chest radiography.The patients were randomly divided into group A (100kv) and group B (70kv).Group B was treated with SAFIRE,and Group A1 and B1with BMI≤ 22 kg/m2,Group A2 and B2 with 22 kg/m2<BMI ≤ 25 kg/m2.Image quality of each group was evaluate subjectively.Evaluate and compare the difference in the image SD,SNR,CNR,CTDI,DLP and ED.Results The image quality of the mediastinum in group B was less than 3 points,which could not meet the need of diagnosis.The noise of A and B group had statistically significant difference (P<0.05),and the noise gradually increased with the decrease of tube voltage.There were significant differencesin the CNR and SNR between the two groups (P<0.05),and the difference of CTDI,DLP and ED between the two groups was statistically significant (P<0.05) Compared with group A,CTDI,DLP and ED decreased by 52 %,48 % and 48% respectively.Conclusion In dual-source CT chest examination,70kV tube voltage combined with iterative reconstruction technique can significantly reduce the radiation dose without affecting the image diagnosis in patients with BMI≤22 kg/m2.