Based on the teeth in vitro skull specimen to investigate CARE kV CT scan technology of decreasing radiation dose

Xian-chun ZENG,Yu-quan WANG,Wei PU,Rui XU,Rong-pin WANG
DOI: https://doi.org/10.3969/j.issn.1006-5725.2016.11.026
2016-01-01
Abstract:Objective Based on the teeth in vitro skull specimen to investigate the feasibility and accuracy of CARE kV CT scan for reducing radiation dose. Methods Of eight fixed and preserved skull teeth were scanned repeatedly in groups of A, B, C. In expose factors of 100 kVp/150 mAs in conventional scanning of group A, open CARE Dose 4D scan of group B and CARE kV scan of group C. We assessed the radiation dose in the group A, B and C, the average CT values of the images, noise, SNR, CNR and subjective ratings. Results Three groups of actual scan tube current, tube voltage and radiation dose were statistically significant (P < 0.05). Compared with group A, the radiation dose in the group B was reduced by about 13.76% and 26.61%reduced in the C group. The average CT values of the images and the noise SNR, CNR and subjective ratings hasd no significant difference (P > 0.05). Conclusion CARE kV technology based on the use specimens tooth can reduce the radiation dose, protect image quality, and be consistent to the reports of previous researches.
What problem does this paper attempt to address?