A study for imaging quality in the course of plain cranial CT scans with different radiation doses

方先来,周全,刘斯润,张建针
DOI: https://doi.org/10.3969/j.issn.1000-0313.2001.03.022
2001-01-01
Abstract:Objective: To evaluate the effect of different radiation doses to imaging quality in the course of plain cranial CT scans.Methods:Eight normal male volunteers were divided into four groups,and each group was performed plain cranial axial CT scans with 110,120,130,140kV and 180 ~ 360mAs respectively. All images were scored at four degrees (excellent, good, poor, very poor) according to grain homogeneity, fine structure appearance of the grey and white matter, hardening of X rays and clearness of tissue interface. Results:The patients were totally scanned 192 times. The image quality with > 120kV and > 240mAs is excellent or good. The rate of superior image quality with ≥ 130kV and≥300mAs, moreover, was higher. Conclusion:Using radiation dose with > 120kV and > 240mAs, the density resolution and SNR may be increased and imaging quality be improved.