Effect of multi-slice spiral CT angiography at different kilovoltage settings on image quality of head and cervical arteries in ischemic cerebrovascular disease

Jun ZHANG,Dao-ying GENG,Xiao-long ZHANG,Rui-jiang ZHU,Hao-xiong LI
DOI: https://doi.org/10.3321/j.issn:1003-3289.2009.03.020
2009-01-01
Abstract:Objective: To investigate the effect of multi-slice spiral CT angiography at different kilovoltage settings on image quality of head and cervical arteries in ischemic cerebrovascular disease. Methods: A total of 70 patients were divided into 2 groups (group A and group B) randomly, and were examined at 100 kV and 120 kV, respectively. The images were reconstructed by maximum intensity projection (MIP) and volume rendering (VR) and displayed by nine-position method. The vascular image quality score was measured and statistical analysis was performed. Results: The vascular image quality score in group A and group B were 2.16±0.36 and 2.83±0.38, respectively. Image quality in group B was better than that in group A, especially the initial part of carotid artery and vertebral artery (P<0.05). The CT dose index in group B was higher than that in group A. Conclusion: The high kilovoltage settings can supply good image quality of multi-slice spiral CTA for head and cervical arteries in ischemic cerebrovascular diseases.
What problem does this paper attempt to address?