Radiation Dose and Image Quality of 70 Kvp Cerebral CT Angiography with Optimized Sinogram-Affirmed Iterative Reconstruction: Comparison with 120 Kvp Cerebral CT Angiography.

Guo Zhong Chen,Long Jiang Zhang,U. Joseph Schoepf,Julian L. Wichmann,Cole M. Milliken,Chang Sheng Zhou,Li Qi,Song Luo,Guang Ming Lu
DOI: https://doi.org/10.1007/s00330-014-3533-y
IF: 7.034
2015-01-01
European Radiology
Abstract:OBJECTIVES:To evaluate radiation dose, image quality, and optimal level of sinogram-affirmed iterative reconstruction (SAFIRE) of cerebral CT angiography (CTA) at 70 kVp. METHODS:One hundred patients were prospectively classified into two groups: Group A (n = 50), 70 kVp cerebral CTA with 5 levels of SAFIRE reconstruction (S1-S5); and Group B (n = 50), 120 kVp with filtered back projection (FBP) reconstruction. CT attenuation values, noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the internal carotid artery (ICA) and middle cerebral artery (MCA) were measured. Subjective image quality was evaluated. Effective dose (ED) was estimated. RESULTS:CT attenuation and noise of the ICA and MCA in Group A were higher than those of Group B (all P < 0.001) while the SNRICA, SNRMCA, CNRICA, and CNRMCA of Group A at S4-5 were comparable to (P > 0.05) or higher than in Group B (P < 0.05). There was no difference in overall image quality between Group A S3-5 and Group B (P > 0.05). ED was 0.2 ± 0.0 mSv for Group A with 85 % ED reduction in comparison to Group B (1.3 ± 0.2 mSv). CONCLUSION:Cerebral CTA at 70 kVp is feasible, allowing for substantial radiation dose reduction. SAFIRE S4 level is recommended for obtaining optimal image quality. KEY POINTS:• 70 kVp cerebral CTA is feasible and provides diagnostic image quality. • 70 kVp cerebral CTA resulted in 85% effective dose reduction. • S4 level of SAFIRE is recommended for 70 kVp cerebral CTA.
What problem does this paper attempt to address?