CT Angiography of the Carotid Arteries with Low Kvp and Low Dose Contrast Media

李玮,刘建新,王霄英,朱颖
DOI: https://doi.org/10.3969/j.issn.1000-0313.2013.05.001
2013-01-01
Abstract:Objective:To evaluate image quality and radiation dose of multidetector computed tomography(MDCT) for carotid CT angiography(CTA) with low kVp and low dose contrast media.Methods:Carotid CTA was performed in 30 patients(group A) with 100kVp and low dose contrast media(0.3mL/kgbw).Carotid CTA was performed in another 30 patients(group B) with 120kVp and 60mL contrast media.Radiation exposure parameters(SSDE and DLP) were compared with independent samples t tests.Image quality was assessed and CT attenuation was measured.Results:Both the mean values of SSDE and DLP of group A were significantly lower than those of group B(24.29±0.98)mGy vs(31.77±2.27)mGy(t=-16.60,P0.01);(889.24±62.00)mGy·cm vs(1185.04±140.05)mGy·cm(t=-10.58,P0.01).There was no significant difference between the overall image qualities of the two groups.Conclusion:Carotid CTA with 100kVp and low dose contrast media can provide satisfactory images with significant reduction of radiation exposure and iodine administration simultaneously.