Feasibility of CT Angiography for the Renal Arteries with Low Dose Contrast Material in the Different Injection Condition

张雷,张追阳,于向荣,王东,高煜,马建勇,华冬英
DOI: https://doi.org/10.13437/j.cnki.jcr.2011.02.042
2011-01-01
Abstract:Objective To investigate the feasibility of 64-slice spiral CT angiography (CTA) with low dose CM for the renal arteries in the different injection condition.Materials and Methods Dynamic CT scan was performed at the same abdominal aorta level in 10 volunteers for measuring the CT value and the theoretical threshold was obtained by the time-density curve (TDC).CTA of the renal artery were performed on 80 patients suspected with renal artery lesions.80 patients were divided randomly into two groups(A and B,with 40 patients in each group)according to the inject method of CM,and then the dose of CM,CT attenuation values of the abdominal aortic artery (AAA) and bilateral renal arteries (RA) were measured for each group and the image quality was also recorded as excellent,good,and poor.Results The dose of group B was less than group A .CT attenuation values of both the AAA and RA were over 200 HU,and no significant difference between two groups was found.The image quality of renal arteries in two groups was well depicted as excellent or good and also was no significant difference.Conclusion It is feasible to use low dose CM in 64-slice spiral CTA for the renal arteries in the different injection condition.