A 400 GΩ Input-Impedance Active Electrode for Non-Contact Capacitively Coupled ECG Acquisition with Large Linear-Input-Range and High CM-Interference-Tolerance

Mingyi Chen,Ho Sung Chun,Ivan Dario Castro,Tom Torfs,Qiuyang Lin,Chris van Hoof,Guoxing Wang,Yong Lian,Nick van Helleputte
DOI: https://doi.org/10.1109/tbcas.2019.2895660
2019-01-01
IEEE Transactions on Biomedical Circuits and Systems
Abstract:Electrocardiogram (ECG) is one of the major physiological vital signs and an effective monitoring method for patients with cardiovascular diseases. However, existing ECG recordings require a galvanic body contact, which is unpractical in daily life. This paper presents the design of an ECG chip that facilitates non-contact ECG recording through capacitive coupling. With the input impedance boosting techniques, as well as an active driven-rightleg (DRL) which boosts common-mode rejection ratio to 70 dB, the single-ended capacitive feedback active electrode (AE) achieves ultra-high input impedance of 400G Omega (<0.5 Hz), a large common-mode interference tolerance (2.8 V-PP), and a high linear-input-range (220 mV(PP)). Implemented in 0.18 mu m 5V CMOS process, the prototype occupies an area of 1.23 mm(2), and consumes 18 mu A and 13 mu A for the AE and DRL, respectively. Real life non-contact capacitively coupled ECG acquisition has been demonstrated, obtaining ECG waves and heart rate in the presence of motion artifacts as well as ambient interference.
What problem does this paper attempt to address?