An Active Dry Electrode Ecg Interface Circuit for Wearable Sensors.

Yintang Yang,Zheng Yang,Zhangming Zhu,Jingmin Wang
DOI: https://doi.org/10.1016/j.mejo.2017.05.008
IF: 1.992
2017-01-01
Microelectronics Journal
Abstract:A high performance active dry electrode interface circuit for ECG signal monitoring in 0.18µm CMOS process is presented in this paper. AC-coupled core amplifier, input impedance boost loop and CMFB (Common-mode Feedback) circuit are applied in this analog front-end circuit. The programmable gain of the core amplifier is set from 12dB to 40dB. With the input impedance boost loop, a constant 2.4GΩ input impedance is achieved (0.1–150Hz @ typical condition). CMFB circuit is utilized to reduce common-mode interference, and the CMRR reaches 85dB with the configuration of ac gain 40dB. In this design, amplifier-based active dry electrode is utilized, which achieves better noise performance than buffer-based active electrode.
What problem does this paper attempt to address?