A High DR High-Input-Impedance Programmable-Gain ECG Acquisition Interface with Non-inverting Continuous Time Sigma-Delta Modulator

Junhao Liang,Sai-Weng Sin,U. Seng-Pan,Franco Maloberti,Rui Paulo Martins,Hanjun Jiang
DOI: https://doi.org/10.1109/A-SSCC47793.2019.9056891
2019-01-01
Abstract:This paper presents a power-efficient, high dynamic range, high input impedance data acquisition circuit for an implantable electrocardiogram (ECG) detector in 65 nm CMOS, occupying an active area of 0.225 mm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . To enhance the dynamic range, we utilize a 3 <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">rd</sup> -order continuous-time sigma-delta modulator with a programmable input gain coefficient and an embedded antialiasing filter. The implementation of the non-inverting integrator allows a high input impedance for the ECG application. The measurement results show that it obtains 84.2dB SNDR in a 150Hz bandwidth with a 1V power supply, consuming 5.4μW at a nominal gain of 0dB. The maximum dynamic range is 99.3dB with the programmable gain activated from 0 to 18dB.