Determination of trace nickel in Bi-based superconductor powder by inductively coupled plasma atomic emission spectroscopy after separation with anion exchange resin and extraction with methylbenzene

Li-xin Fan,Jian-qiang Li,Hui-li Fan,Jian-ling Sun,Xia Zhang,Rui Bao,Qing Lü,Jie Wang
DOI: https://doi.org/10.3964/j.issn.1000-0593(2011)12-3375-04
2011-01-01
Spectroscopy and spectral analysis
Abstract:A new method for the determination of trace nickel in superconductor powder by ICP-AES was proposed. The instrument parameters were optimized, and the matrix effects as well as the method of eliminating interferences were also studied systemically. The results showed that matrix interference in the superconductor powder was serious when the amount of matrix increased, and the repeatability as well as the accuracy was poor, so it was necessary to used separation and preconcentration to improve the accuracy and precision. In this experiment, Ni-diacetyl dioxime complexes were enriched by toluene extraction in alkaline condition after eliminating the interference of major element Bi and partial Pb, Sr, Ca and Cu by eluting with anion exchange resin, then the Ni-complexes were back extracted by diluted hydrochloric acid. The enrichment conditions were investigated in detail and the recovery rate of Ni was higher than 95%. Under the optimal condition, an artificial sample was analysed, the result showed that the found values were identical with reference values, and the RSD and detection limit were 1.9% and 0.19 mu g . g(-1), respectively. The method has been applied to the determination of trace nickel in Bi-based superconductor powder samples with satisfactory results.
What problem does this paper attempt to address?