Determination of Bi in High-Purity Lead and Cadmium Metals by Flame Atomic Absorption Spectrometry after On-Line Matrix Separation

XZ Wu,P Liu,QS Pu,ZX Su
2002-01-01
Abstract:Trace amounts of Bi in high-purity Pb and Cd metals were determined by flame atomic absorption spectrometry in combination with flow injection on-line matrix separation. Matrix separation and Bi(Ill) preconcentration was accomplished by retention of Bi(Ill) on a microcolumn packed with 2-mercaptobenzothiazole modified silica gel at pH 3.0 with a sample flow-rate of 5.6 mL min(-1). Then adsorbed Bi(Ill) was eluted with a mixed eluent containing 3.0 mol L-1 HClO4, 0.5 mol L-1 NaClO4 and 0.2 mol L-1 NaNO3 at a flow-rate of 2.2 mL min(-1) and introduced into the atomizer of the AAS directly. Most of the common coexisting ions did not interfere with the preconcentration and determination. The limit of detection (LOD) of 0.010 mug mL(-1) (which refers to the sample solution), was achieved when 0.3000 mug mL(-1) Bi(III) was preconcentrated for 60 s. The relative standard deviation (RSD) was <2.0% (n = 7) for the proposed method. The linear regression coefficient of the calibration curve was 0.9998 in a working concentration range of 0.10 similar to 2.00 mug mL(-1). The proposed method was successfully applied to the determination of trace Bi in chemically pure Pb grain and Cd band samples.