Development of a Wide-range AO Kelvin Bridge

HUANG Lu,ZHANG Zhong-hua,ZHAO Wei,LI Zheng-kun
2008-01-01
Abstract:For the accurate measurement and calibration of the AC standard resistors in the resistance range from 1 Ω to(10 kΩ),a new type of Kelvin bridge operating at the frequency range from DC to 100 kHz is designed,and a measuring method of substitution is developed.The experimental results of 100 Ω and 1 kΩ standard resistor prove that this bridge takes some advantages,such as high accuracy,simplicity of operating and wide working frequency range,which has already met the specification requirements of the subject of The National AC Resistance Standard.
What problem does this paper attempt to address?