High precision measurement of the load effect of the resistor

Jiang Lan,Zhixin Zhang,Zhengkun Li,Qing He,Bing Han,Shisong Li,Guirong Wang
DOI: https://doi.org/10.1109/CPEM.2012.6250962
2012-01-01
Abstract:The load effect of the resistor should be paid great attention for the precise measurement. Here a high accurate resistance voltage divider (RVD) used to precisely measure the load effect of the resistor is provided. A self-calibrated standard resistor at 100Ω with zero load effect is employed as one arm of the bridge, and several stable resistors with the same temperature coefficient and structure are connected in series as the accurate ratio free from the load effect. In contract to the direct current comparator (DCC), the load effect of the high resistance can be measured due to the low current requirement of the standard resistor. Several experiments have been finished to evaluate the resistance voltage divider and some results are presented here. © 2012 IEEE.
What problem does this paper attempt to address?