Shielding Method for A Resistance Comparison Bridge of Wide Frequency Range

L Huang,ZK Li,ZH Zhang,W Zhao
DOI: https://doi.org/10.3969/j.issn.1001-1390.2006.05.007
2005-01-01
Abstract:Generally, a digital generator and a lock-in amplifier are used as the signal source and the detector for the precise AC resistance comparison bridge of wide frequency range. The measurement results indicate that the obvious disturbance may occur in such a system, which makes the balancing of the bridge difficult. Sometime, the detector cannot be locked in some frequency range. Analyses and theoretic computation show that the problem is from the parasitic coupling between instruments and the ground and the disturbance of the electromagnetic field in space. To solve these problems, an optical isolator, a sine wave generator with very small parasitic capacitance to ground, a metal shielding box and other measures were applied. The uncertainty of the measurement results with the bridge is better than 10(-5) in the frequency range from 40 Hz to 100 kHz.