Low Noise Charge Sensitive Preamplifier A250 for CZT Detector Readout

DENG Zhi,ZHU Wei-bin,LI Yuan-jing,ZHANG Lan,LIU Yi-nong
DOI: https://doi.org/10.3969/j.issn.0258-0934.2007.06.026
2007-01-01
Abstract:Noise performance of a charge sensitive preamplifier-A250 was measured and its widen effect to CZT detector spectrum was analyzed.With 1 s shaping time,A250's zero-capacitance noise is 104.8 e, the noise capacitance ratio is 4.18 e/pF.Energy resolution of 241Am 59.5 keV gamma ray is 4.08%.