Signal Processing Of A Strengthened Electric Field Line Anode Cdznte Detector

Lan Zhang,Jianqiang Fu,Y. Li,Yuanjing Li,Yingshuai Du,Wei Zhang,Jun Li,Yanqing Liu
DOI: https://doi.org/10.1109/NSSMIC.2014.7431293
2014-01-01
Abstract:Most methods to improve the performance of CZT detectors focus on the single polarity charge sensitivity technology, such as Coplanar grid, small pixel effect and so on. Our new design of line anode geometry also applies the single polarity charge sensitivity. With the simple electrode geometry the energy resolution of a 10mm*10mm*10mm CZT detector was improved to 2.17% for 662keV gamma rays. In order to solve the high energy tail and the signal loss of low energy rays incidence from the side wall a revised structure - a Strengthened Electric Field Line Anode was presented. The performance of CZT detectors was improved largely. To obtain better energy resolution kinds of digitized signal processing and analysis methods were tested. The signal peak time difference and the signal amplitude ratio of anode to cathode signals were used to give the relationship with the line anode signal. We could select different value ranges to see the energy resolution variation. Through the cutting of some signals the 662keV gamma ray energy resolution of the 10mm*10mm*10mm CZT was improved further up to about 1.6% without obvious loss of the total energy peak area