Studies of matrix modifiers in the determination of tin by GFAAS

ShuYu Chen,RongHu Wu,Shuqin Lin
DOI: https://doi.org/10.3321/j.issn:0251-0790.1998.06.005
1998-01-01
Abstract:In this paper, the mixed modifiers in the determination of tin by GFAAS were studied. The mechanisms of tin stabilized by two modifiers were inquired using X-ray diffraction analysis. The two modifiers can increase charring temperature up to 1 400 degrees C and 1 350 degrees C, respectively. Mg(NO3)(2) would form a higher background absorption in the. determination process for tin. It has more advantages to use Ni+Vc, such as extremely increasing charring temperature, improving the determination sensitivity, decreasing the background absorption and the matrix interference. The measured results for the stream sediment samples are consistent with the certified value.
What problem does this paper attempt to address?