Analysis on the Adverse Events of Copper-T IUD Breakage of 30 Cases

ZHOU Ding-jie,YAO Jie,WEI Yong-yue,LI Ying
DOI: https://doi.org/10.3969/j.issn.1674-1889.2012.03.017
2012-01-01
Abstract:Objective: To evaluate the adverse events of the copper-T Intrauterine device(IUD)breakage,so as to explore the relative factors.Methods: A total of 30 cases of the breakage adverse events related to copper-T IUD were collected from 105 contraceptive adverse reaction surveillance sentinels from October 2005 to November 2010.All of these cases were analyzed by the method of causality assessment.Results: There were 4 TCu380A cases and 26 TCu220C cases in 30 cases.The fracture parts were cross arms in 28 cases and trailing arms in others.The breakage constituent ratio of TCu220C from B enterprise was higher than that from A enterprise(IC=0.40,lower limit of 95%CI=0.02).Conclusions: The postmarked surveillance of TCu220C from B enterprise should be strengthened.The professional staff of family planning should pay attention to the adverse events of copper-T IUD breakage.To prevent the serious injury events,it is key to keep up the follow-up visit system after insertion.The adverse events should be treated timely,and the serious injury event should be reported immediately.
What problem does this paper attempt to address?