Cupric Ion Release of Different Cu-containing Intrauterine Device in Simulated Uterine Solution

Cao Bianmei,Xi Tingfei
DOI: https://doi.org/10.16153/j.1002-7777.2012.01.007
2012-01-01
Abstract:Objective Cupric ion releases of different Cu-containing intrauterine devices were measured and cupric ion release equations were fit.Methods Cu was determined by flame atomic absorption spectrometry(FAAS).Reproducibility and recovery of the method was examined.Cupric ion release was predicted by Excel regression analysis.Results Time and cupric ion release of TCu220C-IUD show exponential regression relation.Time and cupric ion release of other IUDs show power function regression.The percentage of change for cupric ion release influenced by the change for observed value was from 94.21% to 98.07%.Conclusion The cupric ion releases for different Cu-containing intrauterine device were available by flame atomic absorption spectrometry(FAAS) and Excel regression analysis.The burst release of TCu380A-IUD was the worst.
What problem does this paper attempt to address?