Effective Inhibition of the Early Copper Ion Burst Release with Ultra-Fine Grained Copper and Single Crystal Copper for Intrauterine Device Application

X. X. Xu,F. L. Nie,Y. B. Wang,J. X. Zhang,W. Zheng,L. Li,Y. F. Zheng
DOI: https://doi.org/10.1016/j.actbio.2011.10.018
IF: 10.633
2012-01-01
Acta Biomaterialia
Abstract:To solve the main problems of existing coarse grained copper (CG Cu) intrauterine devices (IUD)—namely burst release and a low transfer efficiency of the cupric ions during usage—ultra-fine grained copper (UFG Cu) and single crystal copper (SC Cu) have been investigated as potential substitutes. Their corrosion properties with CG Cu as a control have been studied in simulated uterine fluid (SUF) under different conditions using electrochemical measurement methods. Long-term immersion of UFG Cu, SC Cu and CG Cu samples in SUF at 37°C have been studied for 300days. A lower copper ion burst release and a higher efficiency release of cupric ions were observed for UFG Cu and SC Cu compared with CG Cu in the first month of immersion and 2months later. The respective corrosion mechanisms for UFG Cu, SC Cu and CG Cu in SUF are proposed. In vitro biocompatibility tests show a better cellular response to UFG Cu and SC Cu than CG Cu. In terms of instantaneous corrosion behavior, long-term corrosion performance and in vitro biocompatibility, the three pure copper materials follow the order: UFG Cu>SC Cu>CG Cu, which indicates that UFG Cu could be the most suitable candidate material for intrauterine devices.