Study on the release behavior of cupric ion in copper-bearing intrauterine devices

Zhang Shuangshuang,Li Ying,Wu Mingjun,Yan Min,Jia Junting,Chen Tong,Liu Jianping
DOI: https://doi.org/10.3969/j.issn.1004-8189.2013.10
2013-01-01
Abstract:Objective: To investigate the release behavior of cupric ion of ten copper-bearing intrauterine devices( Cu-IUDs) immersed in simulated uterine fluid and its influencing factors. Methods: The concentration of cupric ion was determined by flame atomic absorption spectrometer( FAAS). Results: The release behavior of cupric ion of ten Cu-IUDs was varied during the first 60 days after insertion,including burst release pattern such as TCu380A with a fluctuation range from 28μg / d to 4 μg / d and stable release pattern like Yuangong II-300 with a steady release rate at 1-2μg / d. After 60 days,all the IUDs had a stable release rate at 1-2μg / d. IUDs with the same shape had similar releasing curves of cupric ion and the larger the IUDs were,the faster they released. For different shaped IUDs,their release behaviors were different. For example,HCu280 was smaller than Yuangong II-300,but its release rate of cupric ion was much higher than that of the latter. Besides,Yuangong I-220( with 10g indomethacin) released cupric ion faster than Yuangong II-300( without indomethacin),and there was a significant difference between them. Conclusion: The cupric ion release curves are varied mainly during the first 60 days,later,they show no significant difference. IUDs with open shape could release cupric ion easier than those with closed shape. For the same shape,both the increment in area surface of copper and the addition of indomethacin would promote the release of cupric ion.