Development of 4×4 Pixel CdZnTe Detector

ZHANG Lan,LI Yuan-jing,MAO Shao-ji,DENG Zhi,ZHU Wei-bin,LI Shu-wei
DOI: https://doi.org/10.3969/j.issn.0258-0934.2006.03.013
2006-01-01
Abstract:The paper introduced the development of a pixel CdZnTe detector by our research group,including the preparation of the CdZnTe detectors and the performance test.The size of the 4 4 pixel detector is 11 11 6.3mm~(3)and the pixel size is 2 2mm~(2).The energy resolution of the best pixel for ~(137)Cs 662keV gamma ray is 1.4%.After adding a guard ring and the peak position normalization,the total energy resolution is 2.08%.