TAPE HIGH-TEMPERATURE SUPERCONDUCTING RF COILS FOR MRI

He Yan-fa,Wang Yi-nan,Sun Jing,Bai ZHi-ming,Wang Jin-xing
DOI: https://doi.org/10.3969/j.issn.1000-3258.2007.04.009
2007-01-01
Abstract:In order to improve the signal noise ratio (SNR) of low field MRI system, A Bi2223 tape High-Temperature Superconducting RF receiving Coil with detuning circuit has been designed. An electrical coupling mode is used in the coil design and this makes it very suitable for quadrature coil and phase array coil. To prevent degradation by the skin effect, the Ag sheath of Bi2223 tape was firstly removed by using a chemical method, and then the processed tape was wind on a frame to make the final superconducting resonance inductor. A double coil probe method is used to measure the quality factor (Q) of the RF Coils. Results show that the quality factor of HTS RF coils is 2 times that of copper coil with the same structure.
What problem does this paper attempt to address?