Design and simulation of the radio-frequency coils of magnetic resonance imaging

Zhi-ping YAO,Kun YANG,Jian-tao WANG,Qiu-shi REN
DOI: https://doi.org/10.3321/j.issn:1003-3289.2006.09.038
2006-01-01
Abstract:Objective To investigate the simulation and optimization of birdcage coil in the 0.3T Permanent Magnet MRI system. Methods The Biot-Savart's law was applied, a high-pass 8-leg birdcage coil, whose resonant frequency was consistent with ~3H and ~ 129 Xe , was designed through some related calculations and simulations. IDL was used to calculate and simulate our scheme. Results The results showed the uniformity and symmetry of magnetic filed produced by the high-pass 8-leg birdcage coil in a low field. When the legs were reduced to 6 and 4, the uniformity and symmetry fell largely. Conclusion The precision of simulation was good. The uniformity and symmetry of B_1 filed would be damaged when the number of the coil legs decreased.
What problem does this paper attempt to address?