Accurate Evaluation of RF Coil-Tissue Interactions Using a Hybrid FDTD-MoM Method

Wenlong Xu,Feng Liu,Ling Xia,Stuart Crozier
DOI: https://doi.org/10.2529/piers090831185731
2010-01-01
PIERS Online
Abstract:A comprehensive consideration of loading effects is essential to the design/analysis of high-frequency MRI radio frequency (RF) coils. In this study, a hybrid finite difference time domain (FDTD)-method of moments (MoM) simulation method is implemented for the accurate evaluation of RF coil-tissue interaction. Using this approach, the RF coil is modelled using the MoM algorithm, and the biological load is handled with the FDTD algorithm. Effective communication between these two approaches is realized through a Huygence equivalent surface. The algorithm has been successfully validated and applied to a loaded coil study and the simulation demonstrated the potential of the hybrid algorithm for high frequency RF coil designs.
What problem does this paper attempt to address?