On evaluation of MRI coil excited radio-frequency fields inside multilayered spherical head phantom——DyadicGreen's function/MoM method

Ling XIA,Min-hua ZHU,Bin XU,Qing WEI,Feng LIU,Stuart CROZIER
DOI: https://doi.org/10.3785/j.issn.1008-973X.2007.05.035
2007-01-01
Abstract:This work presented a method for computation of magnetic resonance imaging (MRI) coil excited electromagnetic fields (EMFs) inside the human head. A head-sized, multilayered dielectric spherical head phantom with measured radial conductivity and permittivity profiles was constructed, and then DyadicGreen's function (DGF)-based method was used to calculate the EMFs inside the head phantom. This solution is suitable for calculation of EMFs excited by any radio-frequency (RF) resonators whose current distributions can be calculated by method of moment (MoM). The calculated EMFs can then be used to deduce MRI imaging parameters. Results show that the computation time of the proposed method is much lower than that of a full finite difference time domain (FDTD) calculation using a complex head model, and the computation speed is increased by more than ten-fold. Simulation demonstrates that the magnetic and electric fields in human head are affected by the frequency, and that field/tissue interactions increase with the frequency increases, and that the inhomogeneity of the magnetic field in human head increases, thus resulting in the distortion of MR signals. The proposed DGF/MoM method can be used for high field MRI coil design, for example, the rapid evaluation of RF resonator (surface and volume coils) and high performance gradient system designs.
What problem does this paper attempt to address?