SNR-optimized Approach for the Design of RF Surface Coils for Permanent MRI Systems

CHU Xu,Jiang Xiaohua
DOI: https://doi.org/10.3321/j.issn:1000-0054.2005.03.021
2005-01-01
Abstract:The signal-to-noise ratios (SNR) of radio-frequency surface coils with a proper field-of-view were optimized using a design approach based on functional methods. The power losses in the sample and the coil were minimized while the sensitivity at a given point of interest was fixed as specified. Continuous current density distributions were derived and the stream function technique was then used to obtain the conductor paths. A prototype surface coil was designed and constructed for a 0.3 T permanent magnetic resonance imaging (MRI) system. Phantom and in vivo images show that the coil generates a higher SNR than a clinical quadrature body coil at relatively superficial depths and that the SNR exceeds that of a traditional butterfly coil at all depths.