Radiofrequency‐transparent local B0 shimming coils using float traps
Changzhe Liu,Hao Liang,Ming Lu,John C. Gore,Saikat Sengupta,Xinqiang Yan
DOI: https://doi.org/10.1002/mrm.30361
2024-11-06
Magnetic Resonance in Medicine
Abstract:Purpose Static field (B0) inhomogeneities present a major challenge in high‐field MRI. Multicoil shimming using independent, local, direct‐current (DC) shim coils has emerged as a powerful and flexible technique to address this issue. However, many‐turn DC coils can lead to significant mutual coupling with radiofrequency (RF) coils, causing transmit field (B1+) distortions and signal‐to‐noise ratio degradation. Methods We introduce an innovative RF‐transparent DC coil that performs B0 shimming while minimizing RF performance impact. The design incorporates float traps to maintain high RF impedance, allowing flexible placement relative to the RF coil without compromising signal‐to‐noise ratio or affecting B1+. We fabricated square‐shaped DC coils with float traps for 3T MRI and compared them with conventional DC coils. To demonstrate high ΔB0/Amp efficiency, we conducted a B0 shimming experiment around a metal hip implant. Results Bench tests and MRI experimental results demonstrated that the RF‐transparent DC coil effectively minimized RF interference, preserved signal‐to‐noise ratio, and maintained B1+, even when placed near the RF receive coil. Additionally, the DC coil significantly improved B0 homogeneity near metal implants and substantially reduced image distortion. Conclusion The RF‐transparent DC coil offers a flexible, effective solution for managing B0 inhomogeneities, paving the way for integrating multiturn DC coils in clinical MRI settings without extensive hardware modifications.
radiology, nuclear medicine & medical imaging