Low Field NMR System Based on High T_c DC-SQUID

Li Shao,Chu Haifeng,Tian Ye,Ren Yufeng,Chen Yingfei,Chen Genghua,Zheng Dongning
2008-01-01
Rare Metal Materials and Engineering
Abstract:We developed a low field NMR system, which consisting of three parts based on High Tc DC-SQUID. First, the parameters and the relative location of the coils were designed on the basis of thoretical calculation. We made the prepolarization field reached 10mT and the measurement field changed between 200 μT and 1 μT. Second, based on the LabVIEW platform, we designed a programm, by which automatic measurement and data acquisition were achieved perfectly. At last, we fulfilled the test of the SQUID and the system electronics, and all the parts worked well during the process of the comprehensive test. Further experiment and analysis on the system were in process.