Low- T c direct current superconducting quantum interference device magnetometer-based 36-channel magnetocardiography system in a magnetically shielded room

Yang Qiu,Hua Li,Shu-Lin Zhang,Yong-Liang Wang,Xiang-Yan Kong,Chao-Xiang Zhang,Yong-Sheng Zhang,Xiao-Feng Xu,Kang Yang,Xiao-Ming Xie
DOI: https://doi.org/10.1088/1674-1056/24/7/078501
2015-06-25
Chinese Physics B
Abstract:We constructed a 36-channel magnetocardiography (MCG) system based on low-Tc direct current (DC) superconducting quantum interference device (SQUID) magnetometers operated inside a magnetically shielded room (MSR). Weakly damped SQUID magnetometers with large Steward–McCumber parameter βc (βc ≈ 5), which could directly connect to the operational amplifier without any additional feedback circuit, were used to simplify the readout electronics. With a flux-to-voltage transfer coefficient ∂ V/∂ Φ larger than 420 μV/Φ 0, the SQUID magnetometers had a white noise level of about 5.5 fT·Hz−1/2 when operated in MSR. 36 sensing magnetometers and 15 reference magnetometers were employed to realize software gradiometer configurations. The coverage area of the 36 sensing magnetometers is 210×210 mm2. MCG measurements with a high signal-to-noise ratio of 40 dB were done successfully using the developed system.
physics, multidisciplinary
What problem does this paper attempt to address?