Measurements of the Magnetocardiography in Unshielded Environment Using High-Tc Rf SQUID Gradiometer

DF He,F Zhang,FX Xie,YD Dai,SG Wang,P Ma,T Yang,SZ Wang,WC Chen,ZQ Wang
DOI: https://doi.org/10.1016/s0921-4534(00)01429-5
2000-01-01
Abstract:We recorded the 9-point magnetocardiography (MCG) of a normal person in an unshielded environment by using a first-order gradiometer made by subtracting the outputs of two high-Tc YBCO rf SQUID magnetometers. A compensation method was used to cancel the 50Hz line interference. After data processing an averaging, the fine structures of the heart signal (P-wave, Q,R,S-complex, T-wave) could be seen clearly and a SNR of 120 could be obtained.