An Economical Magnetocardiogram System Based on High- T C SQUIDs

Li Zhuo,Zhu Xue-Min,Zhang Li-Hua,Huang Xu-Guang,Ren Yu-Feng,Chen Geng-Hua,Yang Qian-Sheng,Feng Ji
DOI: https://doi.org/10.1088/0256-307x/23/8/097
2006-01-01
Abstract:An economical magnetocardiogram (MCG) system is built in our laboratory. It mainly consists of a MCG data acquisition stage equipped with two high-Tc superconducting quantum interference device (SQUID) magnetometers, a data processing stage with digital filtering and a one-layer μ-metal magnetically shielded room in conjunction with a high-Tc SQUID based active compensation. Experimental results show that a noise level of pico-tesla in MCG profiles, which is necessary for clinical applications, may be achieved with the system. Moreover, stable and convenient operations of the system are demonstrated with simulating MCG measurements.
What problem does this paper attempt to address?