A digital four-arm bridge for the comparison of resistance with capacitance

Yicheng Wang,Stephan Schlamminger
2024-08-23
Abstract:We have built and demonstrated a digital four-arm bridge for the comparison of resistance with capacitance. The digital four-arm bridge mimics the classical quad bridge in the digital domain with three balances: the source balance, the detector balance, as well as the main balance. Due to correlation, the required precision of the source voltages is only of the order of the square root of the ultimate bridge precision. For the comparison of a $100\,$k$\Omega$ resistor with a $1\,$nF capacitor near 1592 Hz, the combined standard uncertainty $(k = 1)$ is $5\times 10^{-9}$.
Instrumentation and Detectors
What problem does this paper attempt to address?
The problem that this paper attempts to solve is to develop a digital four - arm bridge for comparing resistors and capacitors. Specifically, the author constructs and demonstrates this digital four - arm bridge, which simulates the classical four - arm bridge in the digital domain and achieves high - precision impedance comparison through three balance points (source balance, detector balance, and main balance). The main contributions of the paper are as follows: 1. **Improve measurement accuracy**: Through source balance and detector balance implemented by digital technology, these auxiliary balances have only a second - order error on the main bridge balance, thus improving the overall measurement accuracy. 2. **Simplify the measurement system**: Compared with the traditional six - arm bridge, the digital four - arm bridge does not require a complex defining transformer, simplifying the structure of the measurement system. 3. **Low standard uncertainty**: For the comparison of a 100 kΩ resistor and a 1 nF capacitor near 1592 Hz, the combined standard uncertainty (k = 1) reaches 5 × 10^−9, which is comparable to the uncertainty of the NIST traditional four - arm bridge previously reported. ### Specific problem description - **Background**: Traditional four - arm bridges and six - arm bridges have achieved high precision in impedance comparison, but they have some limitations, such as the need for complex defining transformers and high - precision requirements for power supply voltages. - **Objective**: Develop a digital four - arm bridge to use digital technology to achieve high - precision comparison of resistors and capacitors while simplifying the complexity of the measurement system. - **Method**: Achieve source balance and detector balance through digital technology, reduce the requirements for power supply voltage accuracy, and verify its performance through theoretical analysis and experiments. - **Result**: The experimental results show that the digital four - arm bridge has a combined standard uncertainty of 5 × 10^−9 in the comparison of a 100 kΩ resistor and a 1 nF capacitor near 1592 Hz, with high measurement accuracy and stability. ### Mathematical formulas - **Bridge balance condition**: \[ \text{det}(Y)=0 \] where \(Y\) is the admittance matrix of the bridge network. - **Error current**: \[ \Delta I = \rho Y V \] where \(\rho\) is a complex coefficient, \(Y\) is the admittance matrix, and \(V\) is the voltage vector. - **Frequency balance condition**: \[ \omega_0=(R_1 R_2 C_1 C_2)^{-1/2} \] ### Conclusion This paper successfully develops a digital four - arm bridge for high - precision comparison of resistors and capacitors. The source balance and detector balance implemented by digital technology significantly improve the measurement accuracy and simplify the complexity of the measurement system. The experimental results show that the digital four - arm bridge has a combined standard uncertainty of 5 × 10^−9 in the comparison of a 100 kΩ resistor and a 1 nF capacitor near 1592 Hz, and has high application prospects.