Comparison between Thermoelectric and Bolometric Microwave Power Standards

Luciano Brunetti,Luca Oberto,Marco Sellone,Emil Vremera
DOI: https://doi.org/10.1109/TIM.2013.2253914
2021-02-23
Abstract:In the paper, a comparison is described of the microwave power standard based on thermoelectric sensors against an analogous standard based on bolometric sensors. Measurements have been carried out with the classical twin-type microcalorimeter, fitted with N-connector test ports suitable for the frequency band 0.05 - 18 GHz. An appropriate measurand definition is given for being suitable to both standard types. A system accuracy assessment is performed applying the Gaussian error propagation through the mathematical models that interpret the microcalorimeter response in each case. The results highlight advantages and weaknesses of each power standard type.
Instrumentation and Detectors,Classical Physics