Calibration Principle and Correction Factor for Flow Calorimeter in RF Medium-Power Measurement
Jinwen Liu,Wenze Yuan,Yang Zhao,Wei Zhao,Xiaohai Cui
DOI: https://doi.org/10.1109/tim.2023.3306821
IF: 5.6
2023-09-06
IEEE Transactions on Instrumentation and Measurement
Abstract:To establish traceable radio frequency (RF) medium-power calibration and measurement capability, the National Institute of Metrology (NIM), China, designed the flow calorimeter-type RF medium-power standard. Operation frequency and power range of the standard are 10 MHz–8 GHz and 1–100 W, respectively. Subsequently, the calorimeter is operated under the RF condition to calibrate an RF medium-power transfer standard (TS) within the operation frequency and power range. A commercial RF medium-power sensor, as the device under test (DUT), is further calibrated by the TS under identical frequency and power conditions. Hence, the input RF medium power is traced to ac medium power with several correction factors. Though plenty of research on flow calorimeters has been published in the past decades, quantitative evaluation of correction factors in these measurements is still in shortage. In this article, the principle and process of DUT calibration using a flow calorimeter are described. Correction factors due to calorimetric fluctuation, power sensor efficiency of the calorimeter, and dissipated power in cables are analyzed with experiments. Measurement results reflect good stability and repeatability with standard deviation of the mean (SDOM) less than 0.5% of DUT calibration factor, with a mean value of calibration factor of approximately 0.99. Extending on previously published work, in this article, we complete correction factor analysis and the calibration principle with measurement results.
engineering, electrical & electronic,instruments & instrumentation