Impact of MRI RF coil design on the RF-induced heating of medical implants: fixed B1+ rms exposure vs normal operating mode

Aiping Yao,Zihan Li,Mingjuan Ma
DOI: https://doi.org/10.1088/1361-6560/ad2714
IF: 3.5
2024-02-08
Physics in Medicine and Biology
Abstract:A direct comparison of the impact of RF coil design under SAR and B1+ rms limitations are investigated and quantified using RF coils of different geometries and topologies at 64 MHz and 128 MHz. The RF-induced in vivo electric field and power deposition of a 50 cm long pacemaker and 55 cm long deep brain stimulator (DBS) are evaluated within two anatomical models exposed with these RF coils. The associated uncertainty is quantified and analyzed under a fixed B1+ rms incident and normal operating mode. For a fixed B1+ rms incident, the in vivo incident field shows a much higher uncertainty (>5.6 dB) to the RF coil diameter compared to other design parameters (e.g., <2.2 dB for coil length and topology), while the associated uncertainty reduced greatly (e.g., <1.5 dB) under normal operating mode exposure. Similar uncertainties are observed in the power deposition near the pacemaker and DBS electrode. Compared to the normal operating mode, applying a fixed B1+ rms field to the untested implant will lead to a large variation in the induced incident and power deposition of the implant, as a result, a larger safe margin when different coil designs (e.g., coil diameter) are considered.
engineering, biomedical,radiology, nuclear medicine & medical imaging