Impact of Patient Postures on the RF-Induced Heating for the Active Implantable Medical Device (AIMD) at 1.5T MRI
Xiaolin Yang,Ran Guo,Jianfeng Zheng,Xuemin Chen,Ji Chen
DOI: https://doi.org/10.1109/temc.2024.3395983
IF: 2.036
2024-08-18
IEEE Transactions on Electromagnetic Compatibility
Abstract:This article explores the influence of patient postures on radiofrequency induced heating in active implantable medical devices (AIMDs) within a 1.5T magnetic resonance imaging system. Employing the transfer function method, the study assesses RF-induced heating for AIMDs in four different postures. Three AIMD types deep brain stimulation (DBS), cardiac rhythm management (CRM), and sacral neuromodulation (SNM)] are studied across three human models (Duke, Ella, and Fats) at four imaging landmarks under two RF exposure levels: B1+ and specific absorption rate (SAR) restrictions. When comparing temperature rises observed in various postures to those obtained from the typical/standard posture, it was noted that temperature rises can vary significantly. Specifically, temperature rises were found to have variations of up to 82% and 96% for prone and supine arm loop postures, respectively, while remaining within 27% for leg and trunk arm loop postures. Across DBS, CRM, and SNM systems, averaged temperature rises can vary up to 96%, 96%, and 89%, respectively. Noteworthy landmark-specific variations are observed, with changes of up to 43%, 33%, 95%, and 96% for head, cardiac, abdomen, and hip scans. The study underscores the significant impact of patient posture on AIMD heating, emphasizing the necessity for evaluations under diverse conditions.