Potential for high-permittivity materials to reduce local SAR at a pacemaker lead tip during MRI of the head with a body transmit coil at 3 T
Zidan Yu,Xuegang Xin,Christopher M Collins
DOI: https://doi.org/10.1002/mrm.26344
Abstract:Purpose: To illustrate the potential for high permittivity materials to be used in decreasing peak local SAR associated with implants when the imaging region is far from the implant. Methods: We performed numerical simulations of a human subject with a pacemaker in a body-sized birdcage coil driven at 128 MHz with and without a thin (5 mm) shell of material of high electric permittivity around the head. Results: For a shell with relative permittivity of 600, the maximum specific energy absorption rate averaged over any 1 g of tissue near the pacemaker was reduced by 73.5% for a given B1 field strength at the center of the brain. Conclusion: Although further study is required, initial simulations indicate that strategic use of high permittivity materials may broaden the conditions under which patients with certain implants can be imaged safely. Magn Reson Med 78:383-386, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
What problem does this paper attempt to address?