Evaluation of 14 C/ 12 C ratio measurements using accelerator mass spectrometry with standard materials under different graphitization conditions

Sae-Hoon Park,Seung-Won Lee,Yu-Seok Kim
DOI: https://doi.org/10.1016/j.apradiso.2024.111589
IF: 1.787
2024-11-23
Applied Radiation and Isotopes
Abstract:In this study, an evaluation of carbon-beam tuning and 14 C/ 12 C ratio measurements is performed to validate the graphitization process in the pre-processing procedure employed at the accelerator mass spectrometry (AMS) facility of Dongguk University in Korea. The AMS isotopic ratio data are analyzed for samples subjected to three different graphitization conditions: (1) the entire process including Fe catalyst pre-heating with a high vacuum, (2) a high vacuum without pre-heating, and (3) a low vacuum without pre-heating. High-quality isotopic ratio measurements are achieved under these conditions. The measured mean isotopic ratio values for the background samples are almost equivalent to the different graphitization condition samples, with a difference in the order of 10 −16 . Therefore, conducting measurements with or without Fe catalyst pre-heating or high vacuum conditions is equally viable. Furthermore, the simplification of the graphitization process steps reduces the processing time required to produce five samples from 8 to 4 h, thereby enhancing the daily sample throughput by a factor of two, without compromising data quality.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear