A preliminary study of small-mass radiocarbon sample measurement at Xi'an-AMS

Yun-Chong Fu,Wei-Jian Zhou,Hua Du,Peng Cheng,Xiao-Lei Zhao,Qi Liu,Xue-Feng Lu,Wen-Nian Zhao
DOI: https://doi.org/10.1088/1674-1137/39/3/036202
IF: 2.944
2015-03-01
Chinese Physics C
Abstract:To meet the measurement demands on small-mass radiocarbon (carbon content at 10−6 g level) which are becoming increasingly significant, Xi'an-AMS has made improvements to the existing method of sample loading and has upgraded the Cs sputter ion source from the original SO-110 model. In order to study the feasibility of small-mass samples in Xi'an-AMS and evaluate the radiocarbon sample preparation ability using existing routine systems of H2/Fe and Zn/Fe, the small-mass samples prepared by four different methods are tested. They are the mass division method, mass dilution method, H2/Fe reduction method and Zn/Fe reduction method. The results show that carbon mass above 25 μg can be prepared using the existing Zn/Fe system, but no less than 100 μg is required using the existing H2/Fe system, which can be improved. This indicates Xi'an-AMS are now able to analyze small-mass radiocarbon samples.
physics, particles & fields, nuclear
What problem does this paper attempt to address?