NEW SAMPLE PREPARATION LINE FOR RADIOCARBON MEASUREMENTS AT THE GXNU LABORATORY
Hongtao Shen,Junsen Tang,Li Wang,Mingli Qi,Zhaomei Li,Siyu Wei,Kimikazu Sasa,Shulin Shi,Guofeng Zhang,Dingxiong Chen,Linjie Qi,Ning Wang,Houbing Zhou,Ming He,Qingzhang Zhao,Yun He
DOI: https://doi.org/10.1017/rdc.2022.36
IF: 6.324
2022-05-30
Radiocarbon
Abstract:ABSTRACT A new system for preparing 14 C samples was established for a compact accelerator mass spectrometer (GXNU-AMS) at Guangxi Normal University. This sample preparation system consists of three units: a vacuum maintenance unit, a CO 2 purification unit, and a CO 2 reduction unit, all of which were made of quartz glass. A series of radiocarbon ( 14 C) preparation experiments were conducted to verify the reliability of the system. The recovery rate of graphite obtained was more than 80%. The carbon content in the commercial toner and wood sample was linearly fitted to the CO 2 pressure in the measurement unit of the system. The results showed a good linear relationship, indicating that the reliability of the sample preparation system. AMS measurements were conducted on a batch of standard, wood, and dead graphite samples prepared using this system. The results showed that the beam current of 12 C- for each sample was more than 40 μA, the carbon contamination introduced during the sample preparation process was ∼ 2 × 10 –15 , and that the new sample preparation system is compact, low-contamination, and efficient and meets the GXNU-AMS requirements for 14 C samples.
geochemistry & geophysics
What problem does this paper attempt to address?