Sample Preparation System For Carbonate And Dic In Water At The Gxnu-Ams Laboratory

Hongtao Shen,Linjie Qi,He Ouyang,Xinyi Han,Guofeng Zhang,Dingxiong Chen,Kaiyong Wu,Junsen Tang,Li Wang,Ning Wang,Xiaojun Sun,Ming He,Kimikazu Sasa,Jiang Shan
DOI: https://doi.org/10.1017/rdc.2024.65
IF: 6.324
2024-09-19
Radiocarbon
Abstract:A new vacuum line to extract CO 2 from carbonate and dissolved inorganic carbon (DIC) in water was established at Guangxi Normal University. The vacuum line consisted of two main components: a CO 2 bubble circulation region and a CO 2 purification collection region, both of which were made of quartz glass and metal pipelines. To validate its reliability, a series of carbonate samples were prepared using this system. The total recovery rate of CO 2 extraction and graphitization exceeded 80%. Furthermore, the carbon content in calcium carbonate exhibited a linear relationship with the CO 2 pressure within the system, demonstrating its stability and reliability. The system was also employed to prepare and analyze various samples, including calcium carbonate blanks, foraminiferal, shell, groundwater, and subsurface oil-water samples. The accelerator mass spectrometry (AMS) results indicated that the average beam current for 12 C- in the samples exceeded 40 μA. Additionally, the contamination introduced during the liquid sample preparation process was approximately (1.77 ± 0.57) × 10 −14 . Overall, the graphitized preparation system for carbonate and DIC in water exhibited high efficiency and recovery, meeting the requirements for samples dating back to approximately 30,000 years.
geochemistry & geophysics
What problem does this paper attempt to address?